ePubs

The open archive for STFC research publications

Full Record Details

Persistent URL http://purl.org/net/epubs/work/29797
Record Status Checked
Record Id 29797
Title Performance of the cooled pyrolytic graphite analyser bank on the OSIRIS spectrometer at ISIS : Preliminary Results
Contributors
Abstract The pyrolytic graphite (PG) analyser bank on the OSIRIS high-resolution quasi-elastic/low energy inelastic spectrometer at ISIS has been upgraded. The graphite is now cooled close to liquid helium temperature to improve the sensitivity of the spectrometer. In this report we ascertain the performance of the newly upgraded instrument via resolution and intensity measurements collected before and after the modifications. In addition, OSIRIS resolution measurements are compared to resolution data collected from the cooled pyrolytic graphite analyser bank on the IRIS spectrometer.
Organisation CCLRC
Keywords
Funding Information
Related Research Object(s):
Licence Information:
Language English (EN)
Type Details URI(s) Local file(s) Year
Report RAL Technical Reports RAL-TR-2004-005. 2004. raltr-2004005.pdf 2004