Title
Evaluation test of NEXTorr® D 100-5 pump
Contributors
Mr BT Hogan (STFC Daresbury Lab.)
,
Dr OB Malyshev (STFC Daresbury Lab.)
Abstract
A new type of UHV pump, the NEXTorr® D 100-5 pump supplied by SAES getters, was tested in the ASTeC Vacuum Science laboratory. The pump provides the pumping speed characteristics which are similar to a 100 l/s sputter ion pump but required much less space, smaller flange and benefits and a smaller weight. Details of the experimental installation, testing procedure and results are described. The main conclusion is that the pump is very promising for use in the UHV vacuum systems of partial accelerator as well as in the laboratory environment. Long term tests of this pump are still on-going.